Generator Constant: potential, microprocessor-controlled
Working frequency: 145 – 230 KHz with self-adjustment (typically 175 KHz)
Focal spot: 0.4 mm (IEC 336)
Total filtration: 2.0 mm Al @ 70 kV
Anode current: 4 / 8 mA
Voltage at X-ray tube: 60 / 65 / 70 kV (*)
Exposure times: 0.020 – 1.000 seconds, R’10 and R’20 scale
Source-skin distance: 20 and 30 cm
Irradiated field: Ø 55 mm and Ø 60 mm round
Additional collimators: 35 x 45 mm rectangular, 31 x 41 mm and 22 x 35 mm, for sensors size 2 and size 1
Power supply: 50/60 Hz, 115-120 V AC ±10% or 230-240 V AC ±10%
Duty Cycle: Continuous operation with self-adjustment up to 1s/80s total
Arms: (for Standard version only) Available in 3 lengths: 40 cm – 60 cm – 90 cm
Max. arm extension: 230 cm, from wall
Certification: CE 0051, FDA approved
Versions Standard: (wall mounted) or Mobile (on portable cart)